(2010) pp 415-416 A Deep Look at the Effect of Chromogenic Reagent Absorption on the Spectrophotometric MethodSun Feifei School of Science, Tianjin University, Tianjin, 300072;
P.
R. China, sff@tju.edu.cn Published: 22 December 2010 Abstract. Some students have the misconception that monitoring the absorbance of a complex at a wavelength where the free ligand also absorbs strongly willdecrease the accuracy of the spectrophotometric method. This article gives the theoretical basis to show that ligand absorption can be corrected by a reagent blank and clarifies the effect of the undesirable chromogenic reagent absorption.